PT100 Probe  EASY-UP

Temperature probe - 6mm diameter with 3-wire cable


 

Temperature measurement by resistance thermometers relies on a property common to all conductors and semiconductors, which is the variation in their electrical resistance in relation to temperature change. PT100 sensor has a resistance of 100 ohms at 0 °C and the coefficient of variation is  0.00385 ohm per degree centigrade. The technology made to meet modern demands in accuracy is based on thin platinum layers on a ceramic substrate (thin-film resistors).

Since the resistance of the sensor is the total of the resistance of the connecting cables (variable with temperature), a three-wire system is used to cancel out the error.

The PT100 Easy-up probes are assigned a numeric code that once inserted allows the Pixsys controllers to automatically set the main operating parameters.

Ordering codes

2000.90.039
BLI-PT100B-6X50-A304-0000-3 m GSC cable
2000.90.001
BLI-PT100B-6X100-A304-0000-3 m GSC cable
2000.90.153
BLI-PT100B-6X50-A304-0000-3 m TTS cable
2000.90.520
BLI-PT100B-6X80-A304-0000-2 m TTS cable
2000.90.036
BLI-PT100B-6X100-A304-0000-3 m TTS cable
2000.90.521
BLI-PT100B-6X130-A304-0000-2 m TTS cable
2000.90.522
BLI-PT100B-6X230-A304-0000-2 m TTS cable

Sliding junction ordering codes

2000.30.003
Sliding junction Ø 6x1/8" GAS
2000.30.006
Sliding junction Ø 6x1/4" GAS
2000.30.027
Sliding junction Ø 6x1/4" NPT
2000.30.009
Sliding junction Ø 6x1/2" GAS

Main features

Stem diameter
6mm
Immersion material
Steel AISI 304
Internal insulation
MgO Magnesium oxide > 20Mohm at 25°C (500Vdc)
Sensor
PT100 class B (+/- 0,3°C at 25°C)
GSC cable
Silicon rubber, operating temperature -40 ... 250°C
TTS cable
Glass fiber, operating temperature -200 ... 500°C
Sealing
40mm bend proof spring between cable and stem
Response time
6,5 seconds (normative BS 1904/1984/CEI60751)

User Manual - Video - Application Notes - Datasheet

2000.90.039
18/04/2017
English
PDF
2000.90.001
18/04/2017
English
PDF
2000.90.153
18/04/2017
English
PDF
2000.90.520
18/04/2017
English
PDF
2000.90.036
18/04/2017
English
PDF
2000.90.521
18/04/2017
English
PDF
2000.90.522
18/04/2017
English
PDF

Installation

<p>Size and installatioin of temperature sensor pt100 Pixsys srl </p>

Size and installation